Condition
Cause
OCC OUT
• Calibration not yet completed
(repeated)
on the current feeding set
LOCK
• Only the current feeding
schedule is allowed for this patient.
Another feeding program is not
allowed by your healthcare profes-
sional.
ER01 - ER99
• The self test detected an electronic
error
No plug symbol visi-
• The wall outlet doesn't work.
ble, while the pump
is connected to the
• The adaptor is damaged.
mains.

SET UP MODE

The set up mode is used • to set the alarm level,
• to lock the keypad,
• to switch the audible alarm off when a dose is administered,
• to switch the light of the LCD permanently on when connected to the mains,
• to enable connection with a Nurse call or Patient Data
Monitoring System (PDMS).
This set up mode may only be entered by healthcare professionals and authorised personnel trained to use this applica-
tion of the pump.
The patient or his/her relatives without permission of the physician, dietician, nurse or other licensed practitioner, may
not change the settings of the set up mode.
TO SET THE ALARM LEVEL
+
+
ON/OFF
Beep high
&
d
-
Beep low
2 sec
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode.
First the alarm level can be set:
"BEEP HIGH" (+) or "BEEP LOW" (-)
Use the "+" or "-" key to change the setting from "high" to "low" or from "low" to "high".
ZMNL900048_AW.indd 78-79
Correction
• Stop the alarm by pressing the "START/STOP" key.
• Start the pump by pressing the "START/STOP" key and
let it run only briefly.
• Stop the pump by pressing the "START/STOP" key insu-
ring that there has been no occlusion out alarm.
• Remove the feeding set from the pump and reinsert the
feeding set in the pump.
• Restart the pump by pressing the "START/STOP" key.
• The programming feature is blocked in the set up mode
of the pump. Ask your healthcare professional to modify
this setting.
• Turn the pump "off", make sure the pump door is closed
and switch the pump back "on". If the error persists,
contact your healthcare provider or refer the pump to
service.
• Connect the pump to another wall outlet.
• Contact your healthcare professional or
Nutricia subsidiary to replace the adaptor.
-76-
TO LOCK THE KEyPAD
+
Lock
d
FILL SET
-
Unlk
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"UNLK" the keypad (-) or "LOCK" the keypad (+). In the "LOCK" mode the feeding program of the pump
can no longer be changed.
Use the "+" or "-" key to change the setting from "lock" to "unlk" or vice versa.
TO MUTE WHEN DOSE DONE
+
Beep when done
d
FILL SET
-
Mute when done
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"BEEP WHEN DONE" (+): the pump will give an alarm when the dose is administered or "MUTE WHEN
DONE" (-): the pump will not give an alarm after administering the dose. Use the "+" or "-" key to change
the setting from "beep when done" to "mute when done" or vice versa.
TO SWITCH LIGHT ON
+
Lite on
ON/OFF
d
d
FILL SET
-
Lite off
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"LITE ON" (+) the light of the LCD remains on when the pump is connected to the mains. "LITE OFF" (-) the
light of the LCD switches off after a few seconds even if connected to the mains. Use the "+" or "-" key to change
the setting from "lite on" to "lite off" or vice versa.
TO CONNECT WITH NURSE CALL OR PDMS
+
OUTP PDMS
ON/OFF
d
d
FILL SET
-
OUTP NRSE
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"OUTP PDMS" (+) to connect with a Patient Data Monitoring System. "OUTP NRSE" to connect with a NURSE
CALL system. Use the "+" or "-" key to change to setting from "OUTP PDMS", to "OUTP NRSE" or "OUTP
OFF" and vice versa.
Press the "ON/OFF" key and hold it down for 2 seconds to exit the SET UP mode. The settings are auto-
matically saved.