PROGRAMMING THE PUMP
• The Flocare
Infinity
pump can be used for continuous or dose feeding.
®
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessary (see para-
graph: filling the pump set).
• The display shows the total volume delivered since the pump was last cleared.
• If necessary clear the volume delivered by pressing the "CLR" key.
• The last installed flow rate (ml/h) is now displayed.
• The pump is now in the "hold" mode and is ready for programming.
CONTINUOUS FEEDING
+
400 ml/h
d
d
d
ON/OFF
ml/h
START/STOP
-
1 ml/h
• Adjust the flow rate (ml/h) if required by using the "+" or "-" key. Hold either key down to change rapidly.
• If another parameter is indicated in the LCD, it will be necessary to first press the "ml/h" key followed by the
"+" or "-" key to set the flow rate.
• The flow rate ranges from 1 - 400 ml/h in 1ml increments.
• The flow rate slows down and stops shortly at 50 - 125 - 250 ml/h.
• Press the "DOSE=VOL" key and make sure the value is put at 0 ml indicated by the word "CONT". This
means the pump will run in a continuous mode.
• Start the pump by pressing "START/STOP".
• The arcs around the word "run" start rotating.
• The flow rate (ml/h) is displayed whilst functioning.
+
400 ml/h
d
d
d
ON/OFF
ml/h
DOSE=VOL
-
1 ml/h
From the hold mode, with the last flow rate (ml/h) displayed in the LCD:
• Set the flow rate by pressing the "+" or "-" key.
If another parameter is displayed:
• Press first the "ml/h" key followed by "+" or "-" key to adjust the flow rate. The flow rate can be set between
1 and 400 ml/h, with increments of 1ml.
• Press the "DOSE=VOL" key and install the volume to be administered with the "+" and "-" key. The range
goes from 1 - 4000 ml, with steps from 1ml.
• Start the pump by pressing "START/STOP".
During functioning the following parameters can be found in the display:
• By pressing the "ml/h" key the "flow rate" is displayed.
• By pressing the "INFO" key the total volume since the memory was cleared will be
INFO
visualized.
When the required volume or dose has been delivered, "END OF DOSE" will appear and the pump will beep
(or mute, depending upon the pumps' configuration (see paragraph: set up mode).
ZMNL900048_AW.indd 74-75
TO SET A DOSE
+
4000 ml
d
d
START/STOP
-
1 ml
-72-
TO CHANGE THE FEEDING PROGRAM DURING FUNCTIONING
• Press the "START/STOP" key to pause the pump.
• Change the program by pressing the required key (ml/h or DOSE=VOL) and adjust using the "+" or "-" keys.
• Restart the pump by pressing the "START/STOP" key again.
CLEARING THE MEMORy
All parameters and values can be cleared one after the other in the following way:
• Press "START/STOP" to pause the pump.
• Select the parameter that needs to be cleared.
• Press the "CLR" key to clear the memory. The parameter or value returns to its default value:
ml/h = 0 ml/h
VOL = cont
= No dose set. Pump will feed continuously until feed container is empty or pump
is stopped
INFO = 0 ml
= Volume delivered is cleared
In order to keep a clear overview of the daily amount of feed delivered, clear the total volume delivered as
each daily feeding schedule is started, as follows:
• Switch the pump "ON"
START/STOP
• Immediately after the self-test the pump displays total volume delivered.
• Press the "CLR" key. The total volume delivered returns to "0 ml".
• The pump displays the previous programmed flow rate (ml/h).
CLR
• The pump is in the "HOLD" mode ready for programming.
• When another feeding program needs to be started, or the pump will be used for another patient, all set-
tings (rate and dose) can be returned to default and the volume delivered can be set at zero by simply
pressing the "CLR" key and keeping it pressed for 2 seconds.
BATTERy OPERATION
• The Flocare
Infinity
pump is equipped with a Lithium-ion battery.
®
• The battery symbol in the display is automatically replaced by the plug symbol when the adapter is
plugged in. The "fuel gauge" then indicates that the battery is charging by displaying the segments in
an ascending low to high pattern starting with the leftmost segment.
This pattern continuously repeats while the pump is charging.
• To check the status of the battery, disconnect the charger from the pump and turn the
pump on. The bars between E and F (E=Empty, F=Full) represent the "fuel gauge" of the
battery. Each bar is approximately 1/4 of a full battery charge. If 2 bars appear the
battery is half full and remaining operating time is approximately 12 hours at a flow rate
of 125 ml/h.
• In case of power failure the pump automatically switches to battery powered operation.
• When during functioning the last block is gone, the battery symbol will blink to indicate that there is approxima-
tely 1 hour of charge left. The display will flash "BATT" every 3 seconds, alternating with the active display and
the pump will beep every 2 seconds to remind the user of the low battery charge condition. Plug in the charger
to continue to run and recharge the battery.
• As the battery level decreases the pump may not be able to deliver tube feeds at rates greater than 300 ml/h,
even though some battery capacity remains. The pump will alarm and indicate "BATT". If this happens, either