Télécharger Imprimer la page

Philips Respironics SimplyGo Manuel De L'utilisateur page 45

Masquer les pouces Voir aussi pour Respironics SimplyGo:

Publicité

Les langues disponibles

Les langues disponibles

SimplyGo User Manual
Audio Alarm
Outlet Pressure
Expected Service
Life of Device and
Accessories
*Based on an atmospheric pressure of 147 psia (101 kPa) at 70° F (21° C)
** Per ISO 80601-2-69, the maximum Sound Pressure Level is 58 ± 3 dBA and the maximum
Sound Power Level is 66 ± 3 dBA at a continuous setting of 2 LPM The maximum Sound Pressure
Level is 51 ± 2 dBA and the maximum Sound Power Level is 59 ± 2 at Nominal Pulse Mode Setting
of 6 and 20 BPM
SimplyGo Flow Settings and Pulse Volumes
Pulse Mode
10
15
Breath
Rate
15
115
172
20
115
172
25
114
171
30
114
170
35
113
142
40
109
119
+/-15% at STPD*
+/-25% over the rated environmental range
Average of 30 consecutive pulses with 15 m length tubing
42
50 decibels (nominal) at 39 in (1 m)
64 psig maximum (44 kPa)
5 years
Settings
20
25
30
Pulse Volumes (mL)
240
300
360
240
300
360
239
298
356
226
258
289
171
196
220
129
149
168
179
35
40
45
420
480
540
600
420
480
538
595
390
423
440
457
304
318
333
347
233
246
259
271
190
200
209
50
55
60
660
720
655
715
510
562
388
429
303
334
233
257

Publicité

loading