SimplyGo User Manual
Alarm Indicators and Screen Symbols
Note: All SimplyGo alarms are low priority alarms The alarm system should
be veried before use and between users by service personnel in
accordance with the SimplyGo service manual
Technical Alarms
Visual, Audio
Indicators
A sequence of
3 short
beeps
+
1 long beep
that repeats every
4 seconds
22
Description
No Breath Alarm
This alarm occurs
when a breath is not
detected for a period
of 2 minutes or more
It can be silenced by
pressing the Alarm
symbol This alarm
resets itself as soon as
a breath is detected If
no breath is detected
after approximately
30 minutes, the device
shuts down to conserve
power
Note: This alarm is
inactive for ten minutes
after startup Upon
startup, this alarm could
take a maximum of 12
minutes to activate
from the time the alarm
condition is present
Otherwise, it can take
up to 2 minutes to
activate
What To Do
Check the connection
from the cannula to
the device
Make sure the nasal
cannula is properly
positioned on your
face and that you are
breathing through
your nose
Make sure the
cannula tubing is not
kinked or obstructed