Indicators
A sequence of
3 long beeps
+
1 short beep
that repeats every
4 seconds
Description
Low Oxygen
Concentration Alarm
This alarm occurs when
the device is delivering
a lower concentration of
oxygen than specied
This alarm can be
silenced by pressing the
Alarm symbol
Note: This alarm occurs
when the internal O
2
sensor reads < 82% O
This alarm will turn o
if the system is able to
recover from the fault
and achieve >84% O
Note: O
purity is
2
measured at 20 minute
intervals This alarm
could take a maximum
of 31 minutes to
activate from the time
the alarm condition is
present
SimplyGo User Manual
What To Do
Change to another
2
source of oxygen
and contact your
equipment provider