Operating the SimplyGo Device
Power Options
Warning: Do not use the device without the battery installed and operating
To run the device on battery power, make sure the battery is
properly inserted and suciently charged (The battery should be
fully charged the rst time you use it)
To run the device using AC power, rst make sure the battery is
installed Then, follow these steps:
1
Connect the output cable of the AC Power Supply to the
SimplyGo power input connector
2
Connect the AC Power Cord's connector into the AC power
supply
3
Connect the AC Power Cord's plug into an AC outlet installed to
meet national or regional electrical code
Note: The AC Power Supply is equipped with a green light, which
illuminates when it is connected to an energized AC power
source
To run the device using DC power from an automobile or other
motorized vehicle, rst make sure the battery is installed Then, follow
these steps:
1 Start the vehicle
2 Connect one end of the DC Power Cord to the device's power inlet
3 Connect the other end into the vehicle's DC electrical outlet
Note: Because the amount of current available in the vehicle
electrical system is limited, the current available to charge the
SimplyGo battery will be determined by the operating mode
of the device At high settings the time required to charge the
battery will be much longer and at the highest settings, it may
not charge at all
Device Warm-Up
When you turn on your SimplyGo device, it will sense if you are breathing
from it If you are not yet breathing through the cannula, SimplyGo will begin
to pulse automatically about once every ve seconds As soon as you begin
breathing through the cannula, the device will begin delivering pulses based