Caring for your Battery
Handle your battery with care
Do not allow the battery's terminals to touch together as this will
cause it to fail
Do not immerse the battery in water
Do not disassemble or deform the battery
Do not expose to, or dispose of, the battery in re
Avoid exposing the battery to excessive physical shock or vibration
(dropping, etc)
Keep all batteries out of the reach of children
Do not use any battery that has been damaged in any way
Always charge the battery in accordance with the manufacturer's
instructions, using specied chargers only
Do not use modied battery chargers
An optional, external charger/calibrator can be purchased to
recalibrate the battery Contact your home care provider for more
information
Battery Storage
Lithium ion batteries can be stored from -4° F (-20° C) to 140° F (+60° C) at up
to 85% relative humidity However, they are best stored below
70° F (21° C) in a cool, dry, well-ventilated facility free from corrosive gas or
vapor
Storage at temperatures above 113° F (45° C), such as in a hot car, may
degrade battery performance and reduce battery life Storage at low
temperatures may aect initial battery performance