SimplyGo User Manual
Connecting the Optional Humidier
A bubble humidier (sold separately) may be used with the SimplyGo Only
connect the humidier if prescribed Only use the humidier if the device is
in Continuous Flow mode A Humidier Accessory Pouch (REF 1083699) may
be purchased to enable use of a bubble humidier All application accessories
shall include a means to reduce the propagation of re
Warnings
Only use the humidier connector tubing supplied with the
SimplyGo Humidier Connecting Tube kit (REF 1101172) or the
SimplyGo Humidier Pouch and Tube kit (REF 1101602)
Do not overll humidier
Do not reverse the oxygen input and output connections Water from
the humidier bottle will travel through the cannula back to the patient
Do not use the humidier while the device is in Pulse or Sleep mode
To connect the humidier:
1
Place the humidier bottle assembly in the
accessory pouch
2
Remove the cap from the humidier bottle
3
Fill the humidier with distilled water to the level
indicated by the manufacturer
4
Replace the cap and make sure it is tightened securely
5
Attach the humidier bottle adapter tube to the
humidier bottle by turning the wing nut on the bottle
counterclockwise until it is fastened securely (see
Figure 1)
6
Attach the cannula tubing to the humidier bottle
outlet (see Figure 2)
7
Attach the tubing from the humidier bottle assembly
to the oxygen outlet connector on the portable oxygen
concentrator (see Figure 3)
34
Figure 1
Figure 2
Figure 3