SimplyGo User Manual
Visual, Audio
Indicators
1 short beep
that repeats every
4 seconds
A sequence of
3 short beeps
+
1 long beep
that repeats every
4 seconds
26
Description
Low Battery Alarm
This alarm occurs
when approximately
15 minutes of
battery life remains
Remaining battery life
is dependent on your
device settings and your
activity level The empty
battery symbol ashes
on-screen
This alarm can be
silenced by pressing the
Alarm symbol
Note: This alarm could
take up to 45 seconds to
activate from the time
the alarm condition is
present
No Flow Alarm
This alarm occurs when
the device detects there
is no oxygen owing in
the patient cannula
This alarm can be
silenced by pressing
the Alarm symbol
Note: This alarm could
take up to 45 seconds
to activate from
the time the alarm
condition is present
What To Do
Replace the battery
or connect to a power
source
Check the cannula
for kinks or other
obstructions that are
stopping oxygen ow
through it