Curved Probes
The Integrated Probe Positioner (IPP) improves the positioning of approved curved probes in the
trophon2 chamber (refer to the Validated Probe List).
Position the probe curve towards the left of the trophon2 chamber. To engage the IPP, once the
probe is inserted into the probe clamp, push the left side of the IPP and slide to the right (see Figure
13a). This should displace the probe cable till the probe is no longer in contact with the chamber wall
(Figure 12).
To release the IPP, depress the trigger (see Figure 14a) and slide to the left.

C3.4 Closing the Chamber Door

• The chamber door utilises a two stage closure mechanism. Carefully close the chamber door to
the first click and do NOT force shut. The door will automatically close further to seal and lock at
the start of the HLD cycle.
• If the door is not properly closed, a screen message will prompt to close the chamber door.
DO NOT force or manually unlock the chamber door during a HLD cycle.

C3.5 Disinfecting the Probe

Once the door is closed, the trophon2 will request confirmation that the probe is clean and dry prior to
HLD. The door will lock upon confirmation.
1. Scan the trophon AcuTrace Operator Card when prompted (if AcuTrace enabled) and follow the
on screen instructions.
L03072 2.0 04/2020