Error
Effect
No boiling point
Distillation stopped
found
No analogue
The distillation was not started
pump found
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
®
service department,
- send the device for repair, including a short description of the fault.
Operating voltage range
Rated voltage
Frequence
Power input without heating bath
Power input operation "stand by"
Speed
Speed display
Dimensions of visible display area (W x H)
Display
Multiple languages
Right and left movement/ interval operation
Smooth start
Lift
Lifting speed
Stroke
Setting of lower end stop
Head angle adjustable
Timer
Interval
Integrated vacuum controller
Measurement range vacuum
Adjustment range vacuum
Measurement accuracy vacuum
Adjustment accuracy vacuum
Measurement range of differential temperature
Offset balancing range
Interface
Ramp programming
Distillation processes programmable
Remote control
Cooling surface
(Standard condenser RV 10.1,10.10,10.2,10.20)
Minimum cooling water flow rate
Maximum cooling water flow rate
Cooling water pressure
Cooling water temperature range
Fuse
Perm. On- time
Perm. ambient temperature
Perm. humidity
Protection acc. to DIN EN 60529
Protection class
Overvoltage category
Contamination level
Weight (no glassware; no heating bath)
Dimensions (W x D x H)
Operation at a terrestrial altitude
Subject to technical changes!
All manuals and user guides at all-guides.com
Cause
The boiling point could not be
determined
No analogue pump was found

Technical Data

Vac
Vac
Hz
W
W
rpm
mm
mm/s
mm
mm
min
sec
mbar (hPa)
mbar (hPa)
mbar
mbar
K
K
cm
2
l/h
l/h
bar
°C
%
°C
%
kg
mm
m
Correction
- High-boiling solvent: restart
- Check heating bath
- Confirm with ESC
- Check pump and connections
100 - 230 ±10%
100 - 230
50 / 60
100
3,3
5-280
digital
70 x 52
TFT-Display
yes
yes
yes
automatic
50
140
60, contactless
0° - 45°
1-200
1-100
yes
1050 - 1
1012 - 1
+ 2 (+ 1 Digit) if calibritation is done at a constant temperature
adjustable hysteresis
7
±0,5
USB, RS 232
yes
yes
with accessory labworldsoft
®
1500
30
100
1
18 - 22, constantly
2xT1, 6A
250V
5x20
100
5-40
80
IP 20
I
II
2
18,6
500 x 440 x 430
max. 2000 above sea level
75