Electric Slimline Dermatome Models S and S6

Precautions

A thorough, manual cleaning process as detailed in the Cleaning and Sterilization section is
recommended for the Electric Slimline Dermatome Models S and S6. The Dermatome motor
housing cannot be immersed in liquid cleaners, which would occur in an automated process.
Automated cleaning methods may not be effective and may result in damage and reduced
performance of the device.
DO NOT autoclave power supply or power supply wall cord which interconnects the power
supply with the wall receptacle. Damage and reduced performance of the device may result.
EMC Caution: Care should be taken if the equipment is used adjacent to or stacked with other
equipment. If adjacent or stacked use is inevitable, the equipment should be observed to
verify normal operation in the configuration in which it will be used. (Refer to EMC Declaration
Table in the appendix for more detailed electromagnetic emission and immunity information.)
To improve performance and overall safety of the dermatome it is recommended that the
dermatome's power supply wall cord be connected to a Ground Fault Circuit Interrupter
(GFCI) receptacle.
Both the Model S and S6 incorporate sealing devices used to prevent moisture from entering
the handpiece. One or more of these sealing devices can be damaged if disassembled and
reassembled. Refer servicing to an authorized Integra Padgett Dermatome Repair Center.
Use the Model S and Model S6 handpiece with the Model S and Model S6 power supply and
handpiece cable ONLY.
Prior to use, visually inspect the dermatome handpiece and handpiece cable for signs of
physical damage or wear. Do not use the dermatome if damage is observed. The dermatome
handpiece can only be serviced by an Integra Padgett Dermatome Repair Center (refer to
Maintenance and Servicing Section). Dermatome handpiece cable damage may include
cracks or discoloration of the cable jacket, exposed conductors, or distortion of the cable
or connector. If cable damage is observed contact Integra Customer Service immediately to
purchase a replacement handpiece cable (part number 353971513) and discard the damaged
cable.
Used blades should always be disposed of in proper safety containers.
Unit disposal (if necessary) is to be in compliance with your facilities disposal protocols.
Avoid rough handling or dropping the dermatome handpiece and dermatome power supply.
If rough handling is suspected or the unit is dropped, it is recommended to assess the
condition of the dermatome system prior to use as outlined in the Troubleshooting section.