APPENDIX (continued)
Guidance and manufacturer's declaration - electromagnetic immunity
The Integra Padgett Electric Slimline Dermatome Models S and S6 are intended for use in
the electromagnetic environment specified below. The customer or the user of the Integra
Padgett Electric Slimline Dermatome Models S and S6 should assure that they are being
used in such an environment
Immunity Test
IEC 60601 test
level
Electrostatic
±6 kV contact
discharge, (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s) to
line(s)
IEC 61000-4-5
± 2 kV line(s) to
earth
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage
<5 % UT
dips, short
for 0,5 cycle
interruptions
and voltage
<40 % UT
variations on
for 5 cycles
power supply
input lines IEC
<70 % UT
61000-4-11
for 25 cycles
<5 % UT
for 5 sec
NOTE: UT is the a. c. mains voltage prior to application of the test level.
Compliance
Electromagnetic Environment -
level
Guidance
±6 kV
Floors should be wood, concrete
±8 kV
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
±2 kV
Mains power quality should be
that of a typical commercial or
±1 kV
hospital environment.
±1 kV
Mains power quality should be
that of a typical commercial or
±2 kV
hospital environment.
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
><5 % UT
Mains power quality should be
for 0,5 cycle
that of a typical commercial or
hospital environment. If the user
<40 % UT
of the Integra Padgett Electric
for 5 cycles
Slimline Dermatome Models S and
S6 requires continued operation
<70 % UT
during power mains interruptions,
for 25 cycles
it is recommended that the
Integra Padgett Electric Slimline
<5 % UT
Dermatome Models S and S6 be
for 5 sec
powered from an uniterruptible
power supply or a battery.
22