APPENDIX (continued)
Guidance and manufacturer's declaration – electromagnetic immunity
The Integra Padgett
electromagnetic environment specified below. The customer or the user of the Integra Padgett
Slimline
Dermatome Models S and S6 should assure that they are being used in such an environment
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated
IEC 61000-4-3
Foot note test
and foot note test
1
1. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz
to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz
2. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable
RF compliance level above, the equipment should be observed to verify normal operation. If
abnormal performance is ob-served, additional measures may be necessary, such as reorienting
or relocating the equipment
23
®
Electric Slimline
Dermatome Models S and S6 are intended for use in the
IEC 60601
Compliance
test level
150 kHz to
3 Vrms
80 MHz
outside
ISM bands
80 MHz -
3 V/m
2.5 GHz
2
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Integra
Electric Slimline
Padgett
S6 including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2/V1] √P
d=1.2/ √P 80 MHz to 800 MHz
d=2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
the compliance level in each frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
Electric
Dermatome Models S and
, should be less than
1
.
2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Padgett s6

Table des Matières