Electric Slimline Dermatome Models S and S6

Classifications

In accordance with: 60601-1/IEC 60601-1, Type BF
Non-Continuous Operation
Duty Cycle for Dermatome Use: 30 seconds on/5 minutes off
Explosion hazard. DO NOT use in the presence of flammable anesthetics.

Specifications

Maximum Graft Width
Input Voltage
Input Frequency
VA Rating
Output Volts
Output Current
Operating Temperature
Storage and Transport
Assembled Weight
Safety

Warnings

Danger: Explosion hazard: Do not use in the presence of oxygen, nitrous oxide, or other
flammable anesthetics.
These instruments are intended only for professional use in hospitals, physical therapy
locations, and similar medical facilities where the patient will be under the supervision
of trained personnel.
To avoid the risk of electrical shock, this equipment must only be connected to supply
mains with a protective earth.
The power supply contains internally stored energy, and may be a risk for electrical
shock even when unplugged.
5
®
S: 4" (10.16 cm)
100-250Vac
50 / 60Hz
110 VA
48Vac
1A
-20 to 30°C. Humidity 10% to 95% non-condensing, 500 to
1080 hPa
-40 to 55°C. Humidity 10% to 95% non-condensing, 500 to
1080 hPa
S: 2 Lb 5oz. (1.06 kg), S6: 2Lb 12oz. (1.24 kg)
IEC 60601-1, IEC 60601-1-2
Conforms to ANSI/AAMI STD ES60601-1
Certified to CAN/CSA STD C22.2 No. 60601-1
Conforms to Directive 2011/65/EU on the Restriction of the
Use of Certain Hazardous Substances (RoHS II)
S6: 6" (15.24 cm)