Electric Slimline Dermatome Models S and S6
Symbol Definitions (continued)

Troubleshooting:

The following assessment steps can be used to confirm proper dermatome operation. This
assessment is to be performed without the blade and width clip installed.
1.
Visually inspect all dermatome system components for signs of physical damage or
wear including loose parts, part distortion, rough edges, and areas of discoloration.
2.
Check for proper oscillating pin position with the oscillating pin position guide as
outlined below.
3.
Check for proper fit of all dermatome cables. Connect dermatome handpiece cable
between dermatome handpiece and power supply. Connect power supply wall cord to
power supply. All cable connectors are to fit properly to their mating receptacles.
4.
Connect power supply wall cord to wall receptacle and turn power supply switch on.
The indicator light on the power supply is to illuminate. Confirm voltage is available at
the wall receptacle if the indicator light fails to illuminate.
5.
Depress thumb switch on dermatome handpiece momentarily. Dermatome motor is
to run and oscillating pin is to move rapidly from side to side. Note any unusual motor
sounds.
If issues are experienced with any of these system checks, do not use the dermatome and
return the unit to an Integra Padgett Dermatome Service Center for service (see Maintenance
and Servicing section). If no issues are observed during this assessment but performance
issues are experienced, refer to the following troubleshooting table.
PROBLEM
Unit fails to
run with blade
and width clip
installed.
9
®
Atmospheric pressure limitation. The atmospheric pressure values in hPa
units are shown adjacent to the symbol. (Ref. ISO 7000-2621)
Humidity limitation. The relative humidity values are shown adjacent to
the symbol. (Ref. ISO 7000-2620)
POSSIBLE SOLUTION
• The blade and its mating surfaces of the dermatome are to be free of
lint or other foreign objects which may hinder free movement of the
blade.
• Use only the supplied dermatome screwdriver in order to avoid over-
tightening the width clip attachment screws.
• Check condition of width clip for distortion (especially flatness), and
rough edges.