Customer service

The unit may only be repaired by the manufacturer or by an expressly authorised
organisation.
Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.:
+49 (0) 9342/924040
Fax:
+49 (0) 9342/924080
E-mail:
info@uebe.com
Internet:
www.uebe.com
3.

Technical data

Unit type:
Digital automatic unit with electric pump for
measuring blood pressure on upper arm
Dimensions:
L = 115 mm x W = 150 mm x H = 54 mm
Weight:
330 g not including batteries
Display:
LCD display (liquid crystal display) for measured
values and check displays
Memory:
2 memories each with 60 measured values (stored
automatically) and average value (A)
Measurement procedure:
Combines auscultatoric and oscillometric determi-
nation of systole, diastole and pulse
Reference procedure of
Auscultatoric measurement
clinical testing:
Inflating pressure:
approx. 30-40 mmHg above systolic blood
pressure
EN-50
Technical information
Technical information
Pressure display range:
0-300 mmHg
Measurement range
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits:
Blood pressure measurement: corresponds to EN
1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5%
Serial number:
The unit bears a serial number SN which pro-
vides clear identification.
Power supply:
Battery type: 4 x 1.5 V AA alkali manganese (LR
6) or lithium (FR 6) round cells, service life: more
than 800 measurements in 2 years
Optionally stabilised type A1 mains adapter,
output 6 VDC, minimum 600 mA
Cuff with microphone:
Metal ring cuff for arm circumferences of 23-43
cm, type UWK
Operating conditions:
Ambient temperature 10 to 40 °C
Relative humidity up to 85%, non-condensing
Storage and transport
Ambient temperature -5 to +50 °C,
conditions:
Relative humidity up to 85%, non-condensing
Air deflation valve:
Electronically controlled
Automatic switch-off
approx. 3 minutes after end of measurement
Radio clock signal:
DCF77
D
EN-51