Failure en-
Possible cause
countered
Measured
Too little rest before the
values / pulse
measurement. Movement or
values are
talking during measurement.
extremely high
or low
Incorrect measurement
position.
Consecutive
Blood pressure is not a fixed
measurements
quantity. It may vary in the
produce differ-
upward or downward direc-
ent measured
tion within a few minutes by
values (read-
more than 20 mmHg.
ings).
Measurement
Unit detects a fault or that
was interrupted
diastolic value is too low.
and cuff first
Movement during measure-
deflated and
ment
then inflated
No display after
Batteries inserted incorrectly.
unit is switched
Batteries flat.
on
Battery contacts dirty.
No display after
No measured values stored.
Memory button
Batteries flat
is pressed
Measurement
Batteries flat
interrupted
Date/time not
- Function accidentally
shown in display
deactivated
- Function was not reactivated
after battery replacement
EN-50
Technical information

Corrective action

Repeat measurement after
a break of approx. 3-5 mins.
Observe instructions for use
Page 38.
Repeat measurement,
ensuring correct body posture.
(Page 38)
Always take the measure-
ments in the same conditions
in order to obtain comparable
results.
Unit shows no fault. Repeat
measurement.
If necessary, interrupt meas-
urement, measure again after
a break of 5 minutes.
Check position of batteries.
Change batteries.
Clean battery contacts.
Take a new measurement.
Change batteries.
Change batteries.
Activate date and time as
described on page 40.
Technical information

2. Customer service

The unit may only be repaired by the manufacturer or by an expressly authorised
organisation. Please contact:
UEBE Medical GmbH
Service Center
Zum Ottersberg 9
97877 Wertheim, Germany

3. Technical data

Unit type:
Digital automatic unit with electric pump for measuring
blood pressure on upper arm
Dimensions:
L = 122,2 mm x W = 116,5 mm x H = 69,9 mm
Weight:
approx. 214 g not including batteries
Display:
LCD display (liquid crystal display) for measured values
and check displays
Memory:
60 measured values (stored automatically) and average
value (A)
Measurement pro-
Oscillometric determination of systole, diastole and
cedure:
pulse
Reference procedure
Auscultatory measurement
of clinical testing:
Inflating pressure:
Initially up to approx. 180 mmHg, then in stages of 30
mmHg
Pressure display
3-300 mmHg
range:
Measurement range
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits:
Blood pressure measurement: corresponds to EN 1060
Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5 %