13. Using the memory

Measured results are automatically stored in the memory. The memory can store
up to 60 results and the average value. When more than 60 measured values have
been stored, the oldest value (No. 60) is deleted to allow the latest value (No. 1)
to be recorded.

Recalling data

Press the Memory button to call up data. The average
value of the stored results is displayed with "A", the
values for systole, diastole and pulse and the pulse
pressure (PP) are shown alternately in the display.
When the Memory button is pressed again, the value
measured last (No. 1) appears.
When the date/time function is activated, systole, dias-
tole and pulse are displayed alternately with memory
number and date, followed by the pulse pressure (PP)
and the time.
Press the Memory button repeatedly to call up further
measured values.
The stored data remain visible in the display for around
30 seconds. The unit then switches itself off.
EN-46
Operating the unit
Operating the unit

Deleting data

To delete individual values, press the Memory button several times to obtain the
measured value required. Then press the Memory button again and hold it down
for 8-10 seconds until the value first flashes and then disappears.
To delete the entiry memory, recall the relevant average value display (A) and press
the Memory button again, holding it down until the mean value first flashes and
then disappears.
C
EN-47