Failure and error messages

Failure en-
Possible cause
countered
Display Err - 300 Excessive cuff pressure. Cuff infla-
ted to maximum due to movement
of the arm or body.
Display Err - 1
Measurement error. Measurement
could not be carried out due to
movement of the arm or body.
Display Err - 2
Inflation error, pressure built up
too quickly or too slowly.
Fault was recognised.
Display Err - 3
Deflation rate too high or too low. Call the Service Hotline if this
Display Err - 4
Microphone error
Irregular heartbeat, disruptive
movements, shaking, wobbling
(objects), breathing in deeply, etc.
Measured values
Too little rest before the mea-
too high
surement.
EN-48

Technical information

Corrective action
- Repeat measurement
- Do not move arm
- Do not talk
- Repeat measurement
- Do not move arm
- Do not talk
Check that air connector is correctly
fitted, repeat measurement.
- Repeat measurement
- Do not move arm
- Do not talk
occurs repeatedly.
- Insert connector fully into the unit
- Check for correct fitting of the cuff
and the microphone
- If necessary remove any clothing
between the cuff and the upper arm
- Repeat the measurement
If the problem persists, call our
Service Hotline.
Repeat measurement after
3-5 minutes rest. Possible factors
(depending on the severity of the
arrhythmia) affecting the measure-
ment results must be discussed with
your doctor.
Repeat measurement after a break
of approx. 3-5 mins.
- Do not move arm
- Do not talk
Technical information
Failure en-
Possible cause
countered
Unusual measured
Movement or talking during
values
measurement, resting time
not observed, feet possibly
crossed, smoking or drinking
coffee.
Display 0
Start/Stop button accidentally
0
pressed while battery was
being changed.
Measurement was
Unit detects a fault or that
interrupted and
diastolic value is too low.
cuff first deflated
Movement during measure-
and then inflated
ment
No display after
Batteries inserted incorrectly.
unit is switched on
Batteries flat.
Battery contacts dirty.
Dashes in display
No measured values stored.
after Memory
button is pressed
Measurement
Batteries flat
interrupted
Measured values
Incorrect measurement
extremely high
position.
or low
D
Corrective action
Check conditions and repeat
measurement. Observe instructions
for use P. EN-39.
Switch unit off and then on again
with Start/Stop button; if necessa-
ry, activate time signal reception of
radio clock again. See Page EN-40.
Unit shows no fault. Repeat
measurement.
If necessary, interrupt measure-
ment, measure again after a break
of 5 minutes.
Check position of batteries.
Change batteries.
Clean battery compartment.
Take a new measurement.
Change batteries.
Repeat measurement, ensuring
correct body posture. Do not move
your arm and do not talk.
EN-49