1. Failure and error messages

Failure
Possible cause
encountered
Display Err 1
The systolic pressure has been
determined, but the cuff pressure
fell below 20 mmHg afterwards. This
situation occurs if the hose has been
disconnected after having measured
the systolic blood pressure. Further
possible causes: No pulsations could
be detected.
Display Err 2
Unnatural pressure pulse influence
the measurement result. Reason:
The arm has been moved during
measurement (artefact).
Display Err 3
Inflating the measurement takes too
much time. The cuff is positioned
improperly or the hose connection
is leaky.
Display Err 5
The unit detected an unacceptable
difference between systolic and
diastolic pressure. Plausibility check of
the measurement results. Here, values
such as 160/140 are questioned by
the unit.
Display HI
Inflation pressure over 300 mmHg or
pulse over 200, unit can only display
pulse of 40-200, possibly irregular
heartbeat, irregular breathing.
EN-46
Technical information
Corrective action
Large leakage during
measurement. If plug-and-socket
connected ok, send in.
Relax.
If both are not applicable – send
in. Call the Service hotline if this
occurs repeatedly.
Repeat the measurement und
follow the instructions thoroughly.
Please consult the doctor if the
measurement results continue to
be unusual.
Irregular heartbeats - repeat
measurement after 5 minutes
of rest.
Respiratory arrhythmias – repeat
measurement after 5 minutes of
rest, no deep breaths, possibly
consult doctor.
Technical information
Failure
Possible cause
encountered
Display LO
Pulse below 40, unit can only display
pulse between 40 and 200.
Irregular heartbeat, disruptive
movements, shaking, wobbling
(objects), breathing in deeply, etc.
Measured
Was the necessary rest observed
values too
prior to measurement?
high
Has the proper cuff been selected?
Measurement
Upper arm cuff too big.
values too
low
Unusual
Movement or talking during
measured
measurement, resting time not
values
observed, feet possibly crossed,
smoking or coffee cunsumption.
No display
Batteries inserted incorrectly?
after unit is
Batteries flat?
switched on
Battery compartment dirty?
Measurement
Batteries flat?
interrupted
D
Corrective action
Repeat measurement after 3-5
minutes rest. Interrupted pulse
tones? Listen to the peep signal
attentively, possibly consult
doctor.
Repeat measurement after 3-5
minutes rest. Possible factors
(depending on the severity of
the arrhythmia) affecting the
measurement results must be
discussed with your doctor!
Repeat measurement after a
break of approx. 3-5 mins.
- Do not move arm
- Do not talk
Cuffs that are too small result in
excessive blood pressure values.
Measure your arm circumference
in the middle of the upper arm.
Please use standard cuff.
Check arm circumference.
Check conditions and repeat
measurement. Observe
instructions for use P. EN-36.
Check position of batteries.
Change batteries.
Clean battery compartment.
Change batteries.
EN-47