Error
Possible cause
Display LO
Pulse reading is below
40. The device can
only display a pulse of
40–200.
Irregular pulse (cardiac
arrhythmia, movement
during the measure-
ment, tremors, shaking,
deep breathing, etc.)
Readings
Did the patient rest for
are too
the specified amount of
high
time before the meas-
urement was taken?
Abnormal
Movement or talking
readings
while the measurement
is being taken. Patient
did not rest beforehand.

2. Customer service

Device repairs may only be performed by the manufacturer or an
expressly authorised body. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.: +49 (0) 9342/924040
Fax.: +49 (0) 9342/924080
Email: info@uebe.com
Website: www.uebe.com
EN-54
Technical information
Remedy
Repeat the measurement after
resting for 3–5 minutes. Inter-
mittent pulse sounds? Listen
carefully to the beep and check
with your doctor if necessary.
Repeat the measurement after
resting for 3–5 minutes. Always
discuss anything that could
have affected the result (de-
pending on the severity of the
arrhythmia) with your doctor.
Rest for approximately 3–5
minutes, then repeat the meas-
urement.
- Do not move your arm
- Do not speak
Check the conditions (see
"Important usage instructions"
on page 41) and repeat the
measurement.
Technical information

3. Technical data

Model:
visomat
®
Size:
L = 162 mm x W = 110 mm x H = 80 mm ± 2
mm
Weight:
approx. 310 g without batteries
Display:
LCD display (liquid crystal display) for meas-
ured values and check displays
Memory:
2 memories each capable of storing 60
readings
Measurement method: Oscillometric determination of systole, diastole
and pulse
Reference method of
Auscultatory measurement
the clinical test:
Pressure display range: 0 - 299 mmHg
Measurement range:
Blood pressure measurement: 30 - 280 mmHg
Pulse measurement: 40 - 200 pulses/minute
Measurement preci-
Blood pressure measurement (cuff pressure):
sion:
± 3 mmHg, Pulse rate: ± 5%
Power supply:
4 x 1.5 V AA LR6 alkaline manganese batteries,
durability: at least 400 measurements
Optional: visomat
DC, 500 mA
Power consumption:
max. 4 W
Cuff:
Cuff type WXL for an arm circumference of 32 -
52 cm, applied part type BF
Operating conditions:
Ambient temperature 10 °C to 40 °C, relative
humidity 15 % to 85 % non-condensing, air
pressure 700 to 1060 hPa
Conditions for storage
Ambient temperature -20 °C to +50 °C
and transport:
Relative humidity 15 % to 85 % ± 2 %, non-con-
densing
Time to adjust from
2 hours
minimum/maximum
storage temperature to
operating temperature:
Inflation and deflation:
Regulated automatically
comfort XXL
mains adapter, output 6 V
®
EN-55