External Voltage Control
Operating Mode:
Scaling:
Accuracy:
Common mode rejection:
External TTL Control
Operating Mode:
Threshold:
Remote Disable Input
Connection:
Threshold:
PROTECTION
Excess Current:
Excess Power:
Excess Voltage:
Temperature:
Sense Error:
Mode Switching:
General
AC Input:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Cooling:
Safety:
EMC:
Size:
Weight:
Option:
6
The applied voltage determines the operating level within the selected
range.
4 Volts full scale.
± 0·3% ± accuracy of selected range.
Better than –66dB.
The applied signal selects between Level A and Level B settings.
+ 1·5V nominal. A logic high selects Level B.
Terminal block on rear panel.
Apply +3 to +5 V to disable load input.
The unit will trip into the fault state at nominally 92 Amps.
The unit will attempt to limit the power to approx 335 Watts; if this fails
the unit will trip into the fault state at about 360 Watts.
The unit will conduct a current pulse (to absorb inductively generated
spikes) for 1ms at 90V.
The unit will trip into the fault state at nominally 95V.
Surge suppressors will start to conduct above 120V.
The unit will trip into the fault state if the heatsink temperature exceeds
safe levels.
The unit will trip into the fault state if the external voltage sense is more
than 6V below the internal sense.
The unit will trip into the fault state if the mode is changed while the
input is enabled.
110V–120V or 220V–240V AC ±10%, 50/60Hz. Installation Category II.
30VA max.
+ 5ºC to + 40ºC, 20% to 80% RH.
– 40ºC to + 70ºC.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Dual 2-speed fans. Air exit at rear.
Complies with EN61010-1.
Complies with EN61326.
130mm H (3U) x 212mm W (½ rack) x 435mm L.
6 kg.
19-inch rack mount kit.