Audible Alarm And Indicator Lights - AirSep Focus Manuel

Concentrateur d'oxygène portable
Masquer les pouces Voir aussi pour Focus:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
EN
[Read the Important Safety Rules section before operating this equipment]

Audible Alarm and Indicator Lights

When the Focus unit senses inhalation, a pulse of oxygen is delivered through
the nasal cannula. The green light on the unit's control panel blinks each time
a breath is detected.
Additionally, when the unit is operating and the battery is simultaneously being
charged through the Universal AC / DC power supply, the battery for the Focus
unit will display the charge level of the battery (25% to 100% state of charge) on
the battery gauge/indicator(s) and when at 100%, LEDs (lights) remain on for
approximately 15 minutes after reaching a full charge.
An audible alarm sounds if Focus has a low battery, if the cannula is
disconnected, or if performance of the unit is outside of specifications. The light
and audible alarm conditions are explained in detail below and summarized on
the chart later in this section of the manual.
Start-Up
A brief alarm sounds and the green and red LEDs (lights) will alternate at
start-up. Focus begins to operate when the alternating LEDs (light) stop and
green LED (lights) remains on.
Low Battery
Focus unit indicators: As the battery power approaches a low level, the
amber light on the Focus unit will flash on for ½ second with a 5-second
pause and simultaneously, a ½ second alarm sounds with a 5-second
pause. Following the battery warning indicators, if action is not taken, the
unit will shut down. This will be indicated when the amber light flashes
2 times with a 5-second pause and simultaneously, a ½ second alarm
sounds 2 times with a 5 second pause.
Battery indicator: The green light indicator on the battery gauge
(Figure 9) light illuminates intermittently.
When either of these conditions occurs, connect Focus to a DC power
outlet or to an AC power outlet, or change to another source of oxygen
within two minutes. The level of battery charge is indicated by the battery
gauge/indicator(s). You can also check the state of charge at any time by
pressing the button.
As noted above, when the unit is connected to AC or DC power outlet, you
may simultaneously charge the Focus battery supplied with the unit while
using the unit.
EN-26
Focus™
®
AirSep
Corporation
MN172-1 rev. A - 02/13

Publicité

Table des Matières
loading

Table des Matières