Focus™
[Read the Important Safety Rules section before operating this equipment]

Nasal Cannula

A nasal cannula with tubing are used to deliver oxygen from the Focus unit to the
user. The tubing is connected to the unit's oxygen outlet (See Figure 18).
AirSep recommends a nasal cannula with 7 ft (2.1 m) of tubing, AirSep Part No.
CU002-1, or other suitable cannula. Other lengths of non-kinking/star channel
cannula can be used for a total length of up to 25 ft (7.6 m) maximum.
When Focus is operating but does not sense breathing for 15 minutes,
a constant alarm sounds, and the amber alarm light illuminates simultaneously.
If this occurs, check the connection from the cannula to the Focus unit, make
sure that the nasal cannula is positioned properly on your face, and ensure that
you are breathing through your nose. (Your physician may recommend the use of
a chin strap if needed.) If the alarm condition continues, change to another
source of oxygen as available, and contact your Equipment Provider.
®
AirSep