INSTRUCTIONS FOR USE
4.4

Sterilization Instruction:

This device is not intended to be sterilized. Equipment damage may occur.
4.5

Cleaning and Disinfection Instruction:

WARNING:
To help prevent injury and/or equipment damage, obey these warnings:
• Do not reuse wiping materials for multiple steps or on multiple products.
• Cleaning/disinfecting solutions may cause skin rash and/or irritation upon
contact. Follow the manufacturer's instructions found on the product label
and Safety Data Sheet (SDS).
CAUTION:
To help prevent equipment damage, obey these cautions:
• Do not steam clean, submerge, or power wash the Stirrup. Pressure and
excessive moisture can damage the Stirrup.
• Do not use harsh cleansers/detergents, heavy duty grease removers, solvents
such as toluene, xylene, or acetone, and do not use scouring pads (you may
use a soft bristle brush).
• Do not use bleach as your primary everyday cleaner/disinfectant.
RECOMMENDATIONS
These instructions are an example of a cleaning and disinfection process which is proven
to be effective. Follow the applicable facility procedures for reprocessing if they are
proven to be effective.
For proper cleaning and disinfection, staff members should be trained.
The trainer should carefully read the instructions and follow them when the trainee is
being trained. The trainee should:
• Be given time to read the instructions and to ask any questions.
• Clean and disinfect the product while the trainer supervises. During and/or after this
process, the trainer should correct the trainee on any differences from the instructions
for use.
Document Number: 80025141
Version G
Page 41
Issue Date: 26 November 2020
Ref Blank Template: 80025117 Ver. F