SPECIFICATI ONS
DeVilbiss iFill Personal Oxygen Station
Operating Temperature:......................................................................................................................................................................................................41ºF to 95ºF (5ºC to 35ºC)
Operating Humidity Range: .......................................................................................................................................................................................15 to 90% R.H. non-condensing
Operating Atmospheric Pressure Range: .......................................................................................................... 780 hPa to 1013 hPa (2000 meters, ~6500 ft, altitude to Sea level)
Storage Temperature Range: ................................................................................................................................................................................ -13°F to +158°F (-25°C to +70°C)
Storage Humidity Range: .................................................................................................................................................................................................... 5 to 90% non-condensing
Electrical Rating: .......................................................................................................................................................................................... 220-230 V~, 50 Hz, 2.0 A, Single Phase
Operating Voltage Range: .............................................................................................................................................................................................................187 – 253V~ 50 Hz
Power Range: .................................................................................................................................................................................................................................400 Watts Average
Oxygen Purity: ........................................................................................................................................................................................................................................ 93% O
Dimensions: (including casters) without cylinder ............................................................................................................12.25"W x 28.5"H x 22.5"D (31.1 cm x 72.4 cm x 57.4 cm)
Weight: .................................................................................................................................................................................................................................................... 66 lbs (30kg)
Shipping Weight: .................................................................................................................................................................................................................................. 76 lbs (34.5kg)
Ingress Protection: ............................................................................................................................................................................................................................................... IP21
A-Weighted Sound Pressure Level Maximum ..................................................................................................................................................................................................52 dBA
Sound Power Level Maximum ..........................................................................................................................................................................................................................62 dBA
Maximum Limited Pressure normal and single fault conditions .................................................................................................................................................3300 psi (22.75 MPa)
Approvals (50Hz only) EN60601-1 ........................................................................................................................................................................................................................ TUV
Refer to the iFill Service Manual (LT-1929) for Technical Description.

iFill Oxygen Cylinder Typical Fill Times

NOTE – All filling times are approximate and may vary depending on altitude and environmental conditions.
Typical cylinder fill time from empty to 2,000 ± 200 psig (138 ± 13.8 bar) are as follows:
M4 (0.7L) .................60 Minutes
M6 (1.0L) .................75 Minutes
ML6 (1.2L) ...............90 Minutes
NOTE – Degradation of performance may occur if unit is operated outside of specified operating parameters.
WARNING
Auditory alarm signal sound pressure levels that are less than ambient levels can impede operator recognition of alarm conditions.

Description of all auditory alarm signals

The DeVilbiss iFill Personal Oxygen Station includes one alarm condition which indicates Service Required.
The service required alarm is a low priority technical alarm condition.
The alarm is tested at power up, the audible alarm sounds for 0.5 seconds when the iFill is first connected to AC power, this verifies that the audible alert is working.
The device alarms between 4 seconds and 70 seconds after a fault occurs.
The OPERATOR'S POSITION during operation of the DeVilbiss iFill Personal Oxygen Station is normally at a distance or in another room, the iFill operates unattended
while filling oxygen cylinders.
SE-535I
C (1.8L) .............130 Minutes
D (2.9L) .............215 Minutes
E (4.7L) .............350 Minutes