ELECTROMAGNETIC COMPATIBILITY INFORMATION

WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the DeVilbiss iFill
Personal Oxygen Station or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security systems
(metal detectors) as it may cause unacceptable risk to the patient or damage to the DeVilbiss iFill Personal Oxygen Station. Some electromagnetic sources may not be
apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and discontinue use.
Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for
magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
WARNING
Use of accessories and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part
of the DeVilbiss iFill Personal Oxygen Station, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

RETURN AND DISPOSAL

After use of the device, please return the device to the provider for disposal. The equipment, including accessories and internal components, does not belong in your regular
household waste; such equipment is manufactured from high-grade materials and can be recycled and reused. The internal motor, PC boards and wiring should be removed
from the unit and recycled as electronic waste. The metal components can be recycled as metals. Sieve beds can be disposed as household waste. The remaining plastic
components should be recycled as plastics. The European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) requires that electrical and electronic
equipment be collected and disposed of separately from other unsorted municipal waste, with the aim of recycling it. The crossed out waste bin symbol indicates that separate
collection is required.
To minimize the ENVIRONMENTAL IMPACT of the 535 Series Oxygen Filling System during NORMAL USE, remove any oxygen cylinder from the 535 Series Oxygen Filling
System and unplug the 535 Series Oxygen Filling System from AC power. Clean the cabinet air intake filters for the most efficient operation, as listed in the Maintenance
section of this guide.
EN - 12
SE-535I