Technical Specifications; Applied Parts - FLAEM Aspira go P1611EM-20 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22

TECHNICAL SPECIFICATIONS

Mod.
Voltage
Internal battery
supplied:
Battery autonomy:
Device size:
Peso:
Use:

APPLIED PARTS

Type BF applied parts are:
Suction:
Adjustable vacuum
level:
Max air flow:
Noise (at 1 m)
Battery charging time
Ambient temperature
for battery charging
P1611EM-20
12V
14,4V;
2200/2500/2600/2750/
2800/3000/3120 mAh
approximately 50 minutes
36 (W) x 14 (D) x 27 (H) cm
2.5 Kg
30 minutes ON/ 30 minutes
OFF
patient accessories (16)
high vacuum / high flow
from -0.10 a -0.80 bar
(approx)
(accuracy class 2.5)
20 l/min (approx)
63 dB (A) (approx)
minimum 7 hours
From 10°C to 35°C
P1611EM-30
12V
14,4V;
2200/2500/2600/2750/
2800/3000/3120 mAh
approximately 40 minutes
36 (W) x 14 (D) x 27 (H) cm
2.5 Kg
30 minutes ON/ 30 minutes
OFF
patient accessories (16)
high vacuum / high flow
from -0.10 a -0.85 bar
(approx)
(accuracy class 2.5)
30 l/min (approx)
62 dB (A) (approx)
minimum 4 hours
From 10°C to 35°C
17
12 V
VEHICLE POWER
OUTLET POWER CABLE COD.
16667
In the event of a fuse failure,
replace with a 5A - 250V ap-
proved fast fuse, size Ø 6.3 x
30 mm, by undoing the end of
the plug to be inserted in the
vehicle power outlet.
---------------------------------------
POWER SUPPLY MULTI-
VOLTAGE - SWITCHING
COD. 16545
Primary: 100-240V ~ 50/60Hz
Secondary: 12V
5A 60W
---------------------------------------
Operating conditions:
Temperature
min. 0°C; max. 35°C
Relative humidity
min. 10%; max 95%
Storage conditions:
Temperature
min. -5°C; max. 45°C
Relative humidity
min. 10%; max. 95%
Operating/storage
atmospheric pressure:
min. 690 hPa; max. 1060 hPa