TECHNICAL SPECIFICATIONS

Non-lubricated rotary piston suction pump with thermal protection.
Mod.
Supply/power:
Approvals:
Fuse:
Device size:
Weight:
Noise (at 1 m):
Use:
Compliant with Directive:
Performance:
Adjustable vacuum level:
Max air flow:
Applied parts:
Type BF applied parts are:
Operating conditions:
Storage conditions:
Operating/storage
atmospheric pressure:
P1211EM/11 l
230V~ 50Hz 140VA
1 x T 2A 250V
32.5 (W) x 12 (D) x 23.5 (H) cm32.5 (W) x 12 (D) x 23.5 (H) cm 32.5 (W) x 12 (D) x 23.5 (H) cm
2.3 Kg
56 dB (A) (approx.)
Continuous
93/42 EEC
0051
high vacuum /l ow flow
from -0.05 to -0.80 approx.
(accuracy class 2.5)
approx 11 l/min
Possible sterile tube that will be applied to the fitting of the "MANUAL VACUUM FLOW
Temperature min. 5°C; max. 40°C
Air humidity min. 10%; max. 95%
Temperature: min -5°C; max 45°C
Air humidity min. 10%; max. 95%
min. 690 hPa; max. 1060 hPa
P1211EM/13.5 l
230V~ 50Hz 210VA
1 x T 2A 250V
2.4 Kg
56 dB (A) (approx.)
Continuous
93/42 EEC
0051
high vacuum / low flow
from -0.05 to -0.86 approx.
(accuracy class 2.5)
approx 13.5 l/min
CONTROL".
18
P1211EM/20 l
230V~ 50Hz 210VA
1 x T 2A 250V
2.4 Kg
56 dB (A) (approx.)
Continuous
93/42 EEC
0051
high vacuum / high flow
from -0.05 to -0.82 approx.
(accuracy class 2.5)
approx 20 l/min