User Troubleshooting Table

Do Not Ignore Alarms
WARNING: THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EvENT OF A POWER OUT-
AGE, ALARM CONDITION, OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN OR EqUIPMENT PROvIDER FOR THE TYPE OF RESERvE
SYSTEM REqUIRED.
Symptom
The yellow light is on solid.
There is no audible alarm.
The yellow light is flashing. The
alarm is giving 1 beeps every 2
minutes.
The yellow light is flashing. The
alarm is giving 2 beeps every
30 seconds.
The yellow light is flashing. The
alarm is giving 1 beep every 2
minutes.
The yellow light is flashing.
There is no audible alarm.
16 — 20626704-C1 D (English)
Possible Cause
The Eclipse has detected a
problem with the connection
to the power cartridge.
The Eclipse power cartridge is
warm.
The Eclipse power cartridge
voltage is low.
The Eclipse has detected a
flow restriction causing low or
blocked flow.
The Eclipse has detected low
oxygen levels.
Yellow Light
Your Action
1) Remove power cartridge and re-install to ensure that it is secure in the concentrator.
2) If possible, install a separate power cartridge in the Eclipse.
3) If this problem persists, contact healthcare provider.
1) Re-attach the Eclipse to external power or install a fully-charged power cartridge in
the Eclipse. Allow warm power cartridge to cool outside the concentrator for 30 minutes.
2) Re-charge the power cartridge using AC power
3) Replace the power cartridge if another is available.
4) If the condition persists, contact your healthcare provider.
1) Plug Eclipse into AC power supply and recharge or replace power cartridge with fully
charged power cartridge.
1) Clean and replace cabinet inlet filter
2) Ensure that the cannula is not kinked or blocked. If used with a humidifier bottle,
ensure that it is filled properly and not creating a blockage.
3) Ensure that the Eclipse has proper ventilation. It needs to be at least 3 inches from
any surface to ensure the vents aren't blocked.
4) If the problem persists, switch to an alternate source of oxygen and contact health-
care provider for assistance.
1) Ensure the air intake filter is not clogged or restricted. Clean and replace the filter if
necessary.
2) Ensure the Eclipse is in a well ventilated area. Make sure there are at least 3 inches
between the back of the Eclipse and any obstructions (furniture, curtain, etc.)
3) If operating in the car, be sure the back of the Eclipse is facing outward in the seat.
4) If the condition persists, switch to an alternate source of oxygen and contact your
healthcare provider immediately.