NewLife Family
WARNING: USE OF THIS DEVICE AT AN ALTITUDE
ABOVE 3048 METERS (10,000 FEET) OR OUTSIDE
A TEMPERATURE OF 50° F TO 104° F (10° C TO
40° C) OR A RELATIVE HUMIDITY ABOVE 95% IS
EXPECTED TO ADVERSELY AFFECT THE FLOWRATE
AND THE PERCENTAGE OF OXYGEN AND
CONSEQUENTLY THE QUALITY OF THE THERAPY.
WARNING: THE USE OF SOME OXYGEN
ADMINISTRATION ACCESSORIES NOT SPECIFIED
FOR USE WITH THIS OXYGEN CONCENTRATOR
MAY IMPAIR ITS PERFORMANCE.
RECOMMENDED ACCESSORIES ARE
REFERENCED WITHIN THIS MANUAL.
WARNING: TO ENSURE RECEIVING THE
THERAPEUTIC AMOUNT OF OXYGEN DELIVERY
ACCORDING TO YOUR MEDICAL CONDITION
THE NEWLIFE UNIT MUST BE USED WITH
THE SPECIFIC COMBINATION OF PARTS AND
ACCESSORIES THAT ARE IN LINE WITH THE
SPECIFICATION OF THE CONCENTRATOR
MANUFACTURER AND THAT WERE USED WHILE
YOUR SETTINGS WERE DETERMINED.
CAUTION: Federal (USA) law restricts
this device to sale or rental by order
of a physician or other licensed
health care provider.
CAUTION: Do not position the unit so that it is
difficult to access the power cord.
CAUTION: The concentrator should be located
as to avoid smoke, pollutants or fumes.
CAUTION: Ensure concentrator is operated in an
upright position.
CAUTION: Always place oxygen supply tubing
and power cords in a manner that prevents a
trip hazard.
CAUTION: If the audio alarm is weak or does
not sound at all, consult your Equipment
Provider immediately.
CAUTION: The Manufacturer
recommends an alternate source of
supplemental oxygen in the event
of a power outage, alarm condition,
or mechanical failure. Consult your
physician or Equipment Provider for
the type of reserve system required.
8 - ENG
PN MN239-C4 C | User Manual
CAUTION: It is very important to select only the
prescribed level of oxygen. Do not change the
flow selection unless you have been directed to
do so by a licensed clinician.
CAUTION: The Oxygen Concentrator may be
used during sleep under the recommendation
of a licensed clinician. CAUTION: Do not operate
this unit in a restricted or confined space where
ventilation can be limited. This can cause the
device to overheat and affect performance.
CAUTION: Do not allow either the air intake or
the air outlet vents to be blocked. DO NOT drop
or insert any object into any openings on the
device. This can cause the Oxygen Concentrator
to overheat and impair performance.
CAUTION: Operating or storing the Oxygen
Concentrator outside of its normal operating
temperature range can impair the performance
of the unit. Refer to the specification section
of this manual for storage and operating
temperature limits.
CAUTION: Position the unit away from curtains
or drapes, hot air registers or heaters. Be
certain to place the unit on a flat surface and
make sure all sides are at least 1 foot (30 cm)
away from a wall or other obstruction. Do not
place the unit in a confined area. Choose a
dust and smoke free-location away from direct
sunlight. Do not operate the unit outdoors
unless the unit is plugged into a Ground Fault
Circuit Interrupter (GFCI) protected outlet.
CAUTION: In the event of an alarm or you
observe the Oxygen Concentrator is not
working properly; consult the troubleshooting
section of this manual. If you cannot resolve the
problem, consult your Equipment Provider.
CAUTION: If the humidifier bottle tubing is not
properly connected to the humidifier bottle
fitting or to the oxygen outlet, an oxygen leak
can occur.
CAUTION: Normally, you should not need to
adjust the flowmeter on your unit. If you turn
the flowmeter adjustment knob clockwise,
you will decrease and can shut off the flow of
oxygen from your unit. For your convenience,
the flowmeter is marked in ½ LPM increments.
For units equipped with the 2 LPM flowmeter
option, the flowmeter is marked in 1/8 LPM
increments for flow settings up to 2 LPM.