Recommended separation distances between portable and mobile RF
communications equipment and the NewLife Units
The NewLife is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the NewLife can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications equipment (transmitters) and the
NewLife as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Classification
Type of protection against electric shock:
Class II
Protection from electric shock is
achieved by double insulation.
Protective earthing or reliance upon
installation conditions are not
required.
Degree of protection against electric shock:
Type BF Equipment providing a particular
degree of protection against electric
shock regarding
1) allowable leakage current;
2) reliability of protective earth connection
(if present).
Not intended for direct cardiac application.
Degree of protection against harmful ingress of
water:
Drip-proof equipment – IP21.
Protection against ingress of solid foreign
objects greater than 12.5 mm diameter, and
protection against vertically falling drops of
water.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√ P
0.12
0.38
1.2
3.8
12
NewLife Family
(m)
80 MHz to 800 MHz
d = 1.2√ P
0.12
0.38
1.2
3.8
12
Method of cleaning and infection control allowed:
Please refer to Maintenance section in the
NewLife Service Manuals.
Degree of safety of application in the presence of
flammable anesthetic gases:
Equipment not suited for such application.
Mode of operation:
Continuous duty.
PN MN239-C4 C | User Manual
800 MHz to 2,5 GHz
d = 1.2√ P
0.23
0.73
2.3
7.3
23
ENG - 27