Dräger X-am 5600 Notice D'utilisation page 27

Masquer les pouces Voir aussi pour X-am 5600:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
DrägerSensor
Measuring principle
Meauserement value configuration time
t
0...90
Measurement value configuration time t
Measuring range
Lower limit of measuring range 
(EN 45544)
Zero error (EN 45544)
Instrument drift
Warm-up time
Effect of Sensor poisons, hydrogen sulphide
H
S, 10 ppm: halogenated hydrocarbons, heavy
2
metals, substances containing silicone or sul-
phur or polymerisable substances:
Linearity error
Standards
(Measuring function for explosion protection and
measurement of oxygen deficiency and oxygen
surplus as well as toxic gases,
DEKRA EXAM GmbH, Essen, Germany:
BVS 10 ATEX E 080X, PFG 10 G 001 X
7)
Cross sensitivities
1)
The instrument responds to most combustible gases and vapours. The sensitivities differ depending on the type of gas. We recommend a calibration using the target gas to be measured.
2)
Certified for 3 to 500 ppm.
3)
For methane, propane and ethylene.
4)
Certified for 0.01 to 5.00 vol.-% CO
.
2
5)
The measuring signals can be affected additively through acetylene, hydrogen and nitrate monoxide.
6)
For methane, propane and nonane, each with 0-100 %LEL.
7)
A table of the cross sensitivities is contained in the Instructions for Use or the data sheet of the respective sensor.
XXS CO
electrochemical
 25 seconds
6 seconds
0...50
0 to 2000 ppm CO
3 ppm
6 ppm
1 % of the measured
value/month
5 minutes
– – –
3 % of the
measured value
5)
EN 45544-1
EN 45544-2
EN 50271
exist
DUAL IR Ex
IR Ex
infrared
20 seconds for methane
40 seconds for propane
 10 seconds for methane
12 secondsfor propane
0 to 100 %LEL
2)
or
0 to 100 vol.-%
– – –
– – –
– – –
3 minutes
– – –
4 %LEL
EN 60079-29-1
EN 50271
exist
1)
DUAL IR CO
IR CO
1)
infrared
 31 seconds
 15 seconds
0 to 5 vol.-%
3)
0.01 vol.-%
0.02 vol.-%
– – –
3 minutes
– – –
10 % of the
measured value
1) 6)
EN 45544-1
EN 45544-2
EN 50271
exist
2
2
4)
27