Configuration
Only trained personnel are permitted to make changes to the device
configuration.
To individually configure an instrument with standard configuration, the
instrument must be connected to a PC via the USB infrared cable (order no.
83 17 409) or the E-Cal system. Dräger CC Vision PC software is used to
perform the configuration. The Dräger CC Vision PC software can be
downloaded free of charge from the product page of the X-am 5600 at the
following Internet address: www.draeger.com.
Change configuration: see Technical Manual.
Standard instrument configuration:
1)
Dräger X-am 5600
2)
Bump Test Mode
1)3)
Clean air adjustment
4)
Hydrogen added signal
2) 5)
Operating signal
2)
Switch off
2)
LEL factor
4.4 vol. % (4.4 vol. % corresponds to 100 %LEL)
ch
4
4.0 vol. % (4.0 vol. % corresponds to 100 %LEL)
H
2
2) 6) 7)
STEL
Average value duration = 15 minutes
(short-term average)
2) 7) 8)
TWA
(shift average)
9)
can be acknowledged, non-latching, pre-alarm,
Alarm A1
cannot be acknowledged, latching, like main alarm,
9)
Alarm A1 at O
sensor
2
9)
cannot be acknowledged, latching, main alarm,
Alarm A2
®
1)
X-am
is a registered trademark of Dräger.
2)
Different settings can be selected to meet customer requirements on delivery. The current
setting can be checked and changed with the Dräger CC Vision PC software.
3)
Fresh air calibration / zero point adjustment is not supported by the DrägerSensor DUAL IR CO
the DrägerSensor IR CO
or the DrägerSensor XXS O
2
4)
In the case of activated DrägerSensor XXS H
DrägerSensor DUAL IR Ex/CO
(68 11 960) or DrägerSensor IR Ex (68 12 180).
2
5)
A periodic short flashing indicates the operating capacity of the instrument. If there is no
operating signal, correct operation cannot be guaranteed.
6)
STEL: average value of an exposure over a short period, generally 15 minutes.
NOTICE
Extended bump test
on
on
on
allowed
STEL function - disabled
TWA function - disabled
Average value duration = 8 hours
rising flank
falling flank
rising flank
.
3
HC (68 12 025) and activated Ex Channel in the
2
7)
Interpretation only if the sensor is designed for this.
8)
TWA: shift averages are workplace limit values for generally eight hours per day of exposure for
five days a week during a working life.
9)
Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC
Vision PC software.
Activation of the H
offset adds the LEL gas concentration of the activated
2
DrägerSensor XXS H
HC (68 12 025) to the LEL gas concentration of the
2
activated DrägerSensor DUAL IR Ex/CO
DrägerSensor IR Ex (68 12 180) and it is output to the display instead of the IR
Ex display.
Previously set alarm thresholds are retained, so the IR Ex-channel alarm is
triggered earlier in some case in the presence of hydrogen (H
Instrument settings
The following parameters can be changed on an instrument:
Name
Password
1)
Operating signal LED
1)
Horn operating signal
Switch-off mode
2)
Shift length (TWA)
Short-term exposure level (STEL)
4)
1)
At least one of the two operating signals must be switched on.
2)
Corresponds to averaging time and is used to calculate the TWA exposure value.
3)
Interpretation only if the sensor is designed for this.
4)
Corresponds to averaging time and is used to calculate the STEL exposure value.
,
2
(68 11 960) or the activated
2
NOTICE
).
2
Range
Numerical range (3-digit)
Yes/No
Yes/No
"switch-off allowed" or 
"switch-off prohibited" or 
"switch-off prohibited at A2"
60 - 14400 (in minutes)
(Setting for exposure alarm)
3)
0 - 15 (in minutes)
(Setting for exposure alarm)
17