Performance Test And Operating Procedure - Erbe KAUTER N Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

4. Performance test and operating procedure

Inspect the cautery electrodes regularly before application.
Never use cautery electrodes which show any signs of
damage.
1. Insert a cautery electrode into the handle.
2. Connect the unit properly to the AC supply. Activate
3. At a suitable power setting, the electrode tip must glow
If the electrode tip does not glow, in spite of activation:
• Check the power setting.
• Make certain the cautery electrode is firmly attached.
• Check the power connection.
• Inspect the cautery wire for damage.
The unit may only be operated if it functions perfectly
during the test.
WARNING
During operation, the KAUTER N produces heat at the
electrode tip. It must therefore always be ensured that
anesthetics, skin cleansing, degreasing and disinfectant
agents are neither flammable nor explosive. They should
at least have completely evaporated before activating the
unit and be removed from the area of the electrode tip.
Operating procedure
After the unit has been tested as described in the
Performance test chapter, proceed as follows:
1. Insert the power plug into the power outlet. Due to the
2. Insert the plug on the cable (handle) into the socket on
8
the unit.
dark-red.
protective insulation, no connection to protective
ground is necessary.
the front panel.