Erbe KAUTER N Mode D'emploi page 39

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Guidance and manufacturer's declaration - electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
equipment should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted HF IEC
3 V
rms
61000-4-6
150 kHz to 80 MHz
Radiated HF IEC
3 V/m
61000- 4-3
80 MHz to 800 MHz
3 V/m
800 MHz to 2.5 GHz
Compliance level
Electromagnetic
environment - guidance
Portable and mobile HF
communications equipment
should be used no closer to
any part of the equipment,
including cables, than the
recommended separation
distance. The separation
distance is calculated from
various equations depending
on the frequency of the
portable and mobile HF
communications equipment:
Recommended separation
distance
3 V
Equation 1) d=1.2 P
rms
3 V/m
Equation 2) d=1.2 P
3 V/m
Equation 3) d=2.3 P
P is the maximum output
power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer. d is the
recommended separation
distance in metres (m). Field
strengths from fixed
transmitters, as determined by
an electromagnetic site
surveya) should be less than
the compliance level in each
frequency rangeb).
Interference may occur in the
vicinity of equipment marked
with the following symbol:
1/2
1/2
1/2
19

Publicité

Table des Matières
loading

Table des Matières