Wall Mounting

Mounting is usually done on concrete walls or columns, but the unit
can be mounted on any vertical surface. The housing is designed with
spacers on the back to allow moisture to flow behind the housing without
affecting the unit.
Mounting holes are located inside the housing. Open the unit to
access mounting holes.
Drill and mark the holes, as shown:
- Width 11.1 cm (4 3/8") apart (if mounting directly to wall)
- Height 8.3 cm (3.281")
Pre-drill mounting holes from the back of the unit as needed.
Securely mount the unit using the appropriate screws and
anchors. The unit is designed to use #6 screws. Tighten to 8.7
in-lb (1 Nm) maximum.
Close the unit's cover and tighten screws to 29.7 in-lb (3 Nm)
maximum.
Do not remove PC board when removing knockouts.
Knockout on the back of the unit is not for conduit entry.