Routine Maintenance Procedures; Disposal; Cleaning And Storage - Alaris IVAC P6000 Mk II Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Spezifikationen, Caractéristiques, Specifications

Routine Maintenance Procedures

To ensure that this instrument remains in good operating condition, it is important to keep it clean and carry out
the routine maintenance procedures described below. All servicing should only be performed by a qualified
service engineer with reference to the technical service manual (TSM) for this product (Service Manual
reference: 6000PB00001).
Technical service features are available use access code "376", refer to the technical service manual.
If the pump is dropped, damaged, subjected to excessive moisture or high temperature, immediately
take it out of service for examination by a qualified service engineer.
INTERVAL
As Required
12 Monthly

Disposal

The pump should be disposed of taking environmental factors into consideration. To ensure no risk or hazard
remove the internal rechargeable battery and the nicad battery (component B1) from the control board and
dispose of as outlined by the local country regulations. Do not send back to manufacturer. All other components
can be safely disposed of in the normal manner.

Cleaning and Storage

Before the transfer of the pump to a new patient and periodically during the use, clean the pump by wiping over
with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution.
The syringe and extension line are disposable single use items and should be discarded after use according to
their manufacturers instructions.
If the pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged.
Store in a clean, dry atmosphere at room temperature and, if available, employ the original packaging for
protection.
Once every 3 months during storage, carry out functional tests as described in the OPERATIONAL AND ALARM
TEST in the technical service manual and SELF TEST ROUTINE and ensure that the internal battery is fully
charged.
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow liquid to
enter the casing and avoid excess fluid build up on the pump. Do not use aggressive cleaning agents
as these may damage the exterior surface of the pump. Do not steam auotclave, ethylene oxide
sterilise or immerse this pump in any fluid.
82 / 99
Important:
ROUTINE MAINTENANCE PROCEDURE
Thoroughly clean external surfaces of the pump before and after
prolonged period of storage.
1. Inspect AC power supply plug and cable for damage.
2. Perform functional OPERATIONAL AND ALARM TEST. See TSM.
3. Perform the SELF TEST ROUTINE.
4. Perform rate accuracy verification test. See TSM.
5. Perform pressure calibration verification test. See TSM.
6. Operate the pump on battery power until the battery low alarm then
charge the battery to confirm battery operation and charging.
Important:
6001FAOPT71 ISS 8

Publicité

Table des Matières
loading

Table des Matières