The general options available are as follows:
Autosave -
Backoff -
VTBI -
Volume Over Time -
Rate Lock -
Quiet Mode -
AC Fail Alarm -
Rate Titration -
NEOI Warning -
EOI Point -
KVO at EOI -
Beam Alarm -
Purge Rate -
Bolus -
Default Bolus -
Cap Bolus Rate -
Cap Rate -
Display Event Log -
Log Last Patient -
Nurse Call -
Nurse Call Inverted -
Comms Address -
Comms Monitor Only - If not enabled, allows remote control of the unit from the comms links.
Comms Odd Parity -
Comms ASCII -
6001FAOPT71 ISS 8
When disabled this feature clears the patient information on power up.
When enabled this feature requests confirmation to clear or use previous
patient data on power up.
If the backoff feature is enabled it causes the motor to reverse to relieve
line pressure when an occlusion condition occurs.
Enables or disables the VTBI function.
If enabled allows the facility of setting a fixed volume to be delivered over
a fixed period of time.
If enabled the rate can be locked to prevent idle tampering of the set
infusion rate.
If enabled then the low priority alarms and button pushes are muted.
Enables or disables the AC Fail Alarm.
If enabled then the rate can be changed whilst the pump is infusing. To
titrate a rate while infusing, use the rate adjust keys to select the new rate
and press the START button to confirm the new rate.
Sets the NEOI warning in time. 1 to 15 minutes.
Sets the end of infusion volume as a percentage of the syringe volume.
Pump runs at KVO for half this volume. 0.5% - 5%
If enabled the pump will switch to run at the KVO rate 1.0ml/h (or the
current set rate if this is lower) at the EOI point. If disabled the pump will
stop at the EOI point.
Sets up the default beam level. Select from 8 levels - 0 to 7.
Sets the purge rate.The upper limit prevents excessive loss of fluid whilst
purging.
Bolus enabled.
Sets the default bolus rate. 10 - 1200ml/hr.
Sets maximum permissible value for the bolus rate. 10 - 1200ml/hr.
Sets maximum permissible value for the infusion rate. 1 - 1200ml/hr.
If enabled the event log can be accessed via the options menu.
Event log shows all patients, or just the last (since new patient selected).
Enables or disables the nursecall option.
When enabled the Nurse Call Option output function is inverted.
Sets the Pumps Communications Address.
Monitor is always possible.
If enabled, communication parity is odd.
If enabled, communications protocol is in ASCII.