Alaris IVAC P6000 Mk II Mode D'emploi page 22

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Getting Started
Alarm Procedures
Description
Display
BATTERY EMPTY - Internal battery empty. To silence
BATTERY EMPTY
the alarm switch the power switch to the OFF position
and reconnect the pump to AC power. Restart operation
on AC power to charge the internal battery.
NEAR END OF SYRINGE WARNING (NEOI) - This is a
NEAR END OF SYRINGE
timed event and can be set from 1 to 15 minutes. To set
/ change this value see the general options section in
this user manual.
SYRINGE EMPTY - END OF INFUSION (EOI) - The
pump has reached the end of the infusion. About 1%
(can be set in configuration) of the syringe volume will
SYRINGE EMPTY
remain in the syringe to minimise the risk of the infusion
of air bubbles into the set. If KVO at EOI is enabled the
rate will drop to the keep vein open (KVO) rate.
Approximately half the remaining volume in the syringe
will be delivered at the KVO rate.
AC POWER SUPPLY DISCONNECTED WARNING AC
Power has been disconnected and the pump is
AC POWER FAIL
operating on battery power. Reconnect AC power or
press START to silence the alarm and continue battery
operation. The display will flash AC POWER FAIL. The
alarm will automatically cancel if the AC power is
reconnected.
INTERNAL MALFUNCTION - The alarm system has
ERROR
detected an internal malfunction. Note the malfunction
code. Remove the unit from service for examination by
a qualified service engineer.
BATTERY WAS LOW AT
BATTERY MAY BE LOW/EMPTY - When the pump was
LAST POWER OFF. HAS
last turned off the battery was low. If the pump has not
been charged since then operate on AC power only.
PUMP BEEN RECHARGED
Alarm Procedures without display messages
Description
NURSE ATTENTION WARNING - Unit left switched
ON for over 2 minutes without starting operation. Press
3 BEEPS
STOP or any of the control buttons to silence the alarm
for a further 2 minutes. Alternatively, press and hold
down STOP for 2 seconds to delay call back for 15
minutes.
16 / 99
6001FAOPT71 ISS 8

Publicité

Table des Matières
loading