Configured Options
Enable Syringes
Enable syringes is used to pre-configure the type and size of syringe allowed for use by an
operator. Select all the syringes which may be used with the pump. Disable any syringes which
will not be used.
1. Press the ON/OFF power button to the
OFF position
2. Hold down the START button and press
the ON/OFF button.
3. The main display will show "000". Enter the
Access Code "251" using the rate selection
arrows.
4. Enter the first digit and press NEXT
indicated on the screen.
5. When the complete code has been entered
press the OK softkey indicated on the
screen.
6. Select ENABLE SYRINGES using the
chevron keys and press the OK softkey
indicated on the screen.
7. Use the chevron to toggle up and down the
list of available syringe types and sizes.
Press
the
MODIFY
enable/disable the syringe.
8. Press the OK softkey to store the syringes
enabled.
9. When the set-up is complete, turn the
pump OFF and return the pump into
service,
or
configuration option.
22 / 99
softkey
select
the
next
to
6001FAOPT71 ISS 8