Getting Started

Pressure Level

1. To check and adjust the pressure level
press the PRESSURE button. The display
will change and a bar graph will show the
pressure alarm level and current pressure
level.
2. To adjust the alarm level press the rate
keys to increase or decrease the level. The
new level will be indicated on the display.
3. When the pressure level has been selected
press QUIT to exit the screen.

Volume To Be Infused (VTBI)

VTBI
This option allows you to set a specific volume to infuse, and rate at the end of the volume to be
infused, from stop, 1ml/h, 2ml/h and continuous infusion at the set rate.
1. Press the VTBI softkey to select the VTBI
option.
2. Enter the volume to be infused using the
chevron keys.
3. When the volume to be infused has been
entered, press the OK softkey.
4. Select the rate at the end of the VTBI using
the rate selection chevrons. The default is
STOP.
5. Press the OK softkey to exit the VTBI
menu.
6. When the pump has delivered the set
volume it will alarm. Press the CANCEL
softkey to clear the alarm.
7. Press the CLEAR softkey to turn the VTBI
function off, or set a new VTBI using the
chevron keys.
10 / 99