Hach LANGE EVITA OXY Manuel D'utilisation page 96

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 165
the container in order to induce a potential equalization between medium
and PE. If the flow rate in the container is high, a static potential may
occur locally. Place in this case the ground electrode near to the
transmitter.
Startup
Transmitter OXY 4100/4150/3150 without USC 5000/
7000 signal converter
Figure 16 Transmitter without signal converter
1.
The power supply is connected whereby the OXY 1100 sensor
points upward in the open air for two minutes. This generates a
zero point adjustment of the service life counter of the sensor. The
data of the service life counter can be read at the signal converter
(USC) or through HART
®
2.
Place the transmitter such that the sensor OXY 1100 points
downward outside of the medium. Leave it for 1 hour in this position
so that the sensor can stabilize.
3.
Calibrate the sensor in accordance with instructions
Calibration).
English 96
communication.
(section
System with USC 5000/7000
Figure 17 System with USC 5000/7000
1.
Place the transmitter such that the sensor OXY 1100 points
downward outside of the medium (refer to
2.
Switch on the signal converter USC 5000/7000.
3.
Leave the transmitter for at least 1 hour in this position so that the
sensor can stabilize before it is calibrated.
4.
Check the default settings and apply changes if necessary.
5.
Reset the operating hours counter.
6.
Calibrate the sensor in accordance with instructions
Calibration).
7.
When the calibration process is complete, insert the transmitter into
the medium to be measured.
Calibration, page
106).
(section

Publicité

Table des Matières
loading

Table des Matières