Monitoring of the current- and of the relay outputs
Figure 35 Monitoring of the current- and of the relay outputs
Service mode
In the service mode there is the option to monitor the relay
and the current outputs directly. This can be very helpful if
for example a control loop should be
set or an alarm signal should be checked.
The quality of the HART® communication is shown in the
sub menu: "HART comm. Error".
The quality is shown as follows:
VVVV XXXX YYY % ZZ
where:
VVVV is the number of the reports without error,
XXXX is the number of the reports with error,
YYY % is messages with errors in percent,
ZZ is the number of the reports that have been deleted
because for example the maximum number of repetitions
has been reached.
English 112
USC
USC
USC
Service
I Out 1
I Out 2
mode
I OUT 1
Mode
Normal
Normal
I OUT 1
Forced
Forced value
00.0 mA
00.0-24.0 mA
USC
USC
Relay 1
Relay 2
RELAY 1
Mode
Normal
Normal
Open
Closed
USC
USC
Digital input
Event log
High
(High/Low)
USC
Log xxxxdxxh
xxxxxxxxxxxxxxxx
USC
USC
I Out 3
I Out 4
USC
USC
USC
Relay 3
Relay 4
Relay 5
USC
HART comm. error
0856 0007 001% 00