Hach LANGE EVITA OXY Manuel D'utilisation page 114

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 165
Figure 37 Error codes
E
O
5
5
EO55: The membrane of the OXY 1100
sensor is damaged.
English 114
For each event there is a unique
identification code that consists of 4 digits.
1st digit:
Category
I (Information);
W (Warning);
E (Error)
2nd digit:
Sensor type
U for USC;
O for OXY
I for INSITU 4100;
N for INSITU 5100;
P for PH/REDOX
S for TSS
3rd and 4th
Error code 00–99
digit:
These codes
denote the following
events:
0–9 Information
10–49 Warning
50–99 Error
Information
The system proceeds as normal with the measurement procedure. Relays and
output signals are not affected.
Warning
The system proceeds with the measurement procedure, however an event has
occurred that could impair the function of the system and that makes the
intervention of an operator necessary.
The reason for this warning may however disappear again by itself.
If the warning relays have been set, the warnings are shown in accordance with
the setup.
Error
The entire system or a part of the system is malfunctioning and the output
signal no longer works reliably. The faults can only be rectified by a technician.
If the error relays have been set, the errors are shown in accordance with the
setup.
®
The error list of the EVITA
OXY transmitter shows the 5 most recently
occurred errors that are still pending, and the event log shows the last
25 events, in other words data, warning- and error messages.
The error list of the USC signal converter can show up to 14 pending error- and
warning messages, whereby first the USC errors and then the errors of the
transmitter are shown.
With more than 14 errors the message "Additional errors cannot be viewed" is
shown on the display. The USC event log shows the last 9 events.

Publicité

Table des Matières
loading

Table des Matières