Symptom
Current output
Flashing bell symbol
Not defined
in the display
Event
Cause
code
USC cannot read the data for current output
I OUT 1
1
No response from the transmitter
2
Transmitter has received a new address
3
Another master is calling up problems with the
®
EU61
HART
communication.
4
Supply voltage for transmitter
too low
5
Too much noise interference on the current loop
6
Transmitter defective
7
USC defective
USC cannot read the data for current output I OUT 2
EU62
1–7: As with EU61
USC cannot read the data for current output I OUT 3
EU63
1–7: As with EU61
USC cannot read the data for current output I OUT 4
EU64
1–7: As with EU61
USC cannot read the data for relay 1
EU65
As with EU61
USC cannot read the data for relay 2
EU66
As with EU61
USC cannot read the data for relay 3
EU67
As with EU61
USC cannot read the data for relay 4
EU68
As with EU61
USC cannot read the data for relay 5
EU69
As with EU61
Resolution
1
Check cable and shielding
2
Switch USC off and on again
3
Disconnect other master, or set up
USC for other master, refer to figure
51, page 52
4
Check supply voltage of the transmit-
ter; it must be at least 20 V.
5
Check cable and shielding
6
Replace transmitter
7
Replace USC
As with
EU61, page 117