3. To start the particle count, push
button while the count is measured.
Note: Push
to end the test before the count is complete. Incomplete particle
count data will not be stored or printed unless the Save Partial Data option is
selected in System Settings.
4. When the count measurement is complete, the test will stop
automatically.
Change the particle count location
There are two methods to change the location for a particle count.
• On the Sample screen, push the location name. Select the new
location name and confirm.
• On the Sample screen, push the PLUS button to increment the
location, or push the MINUS button to decrement the location.
See settings during the particle count
Location and group settings can be seen at any time during the particle
count cycle.
• On the Sample screen, select the Settings tab on the right side of the
screen.
See historical data during the particle count
Historical sample data can be seen at any time during the particle count
cycle.
1. On the Sample screen, select the ARROW button.
2. Select the HISTORICAL icon to see the data.
Use the filter scan probe
N O T I C E
The filter scan probe function applies to 1 CFM units only.
1. On the Counter Navigation screen, push SAMPLE.
2. In the Test screen, push
.
3. To start the test, push START FILTER PROBE TEST.
. The
icon will change to a
4. To end the test, push STOP FILTER PROBE TEST.
5. Push
to generate a brief report of the last completed test.
How to use the Print Center
About the Print Center
N O T I C E
To avoid damage to the printer, do not operate the printer without paper. If the
particle counter must be used without paper, be sure to set the Sample Print
Mode field to "None".
The particle counter has a built-in printer. The Print Center screen is
accessible from the:
• Counter Navigation screen
• Historical screen
• System Diagnostics screen
• Test/Report Wizard screen
• Area/Location Setup screen
On the Print Center screen, the user can:
• Set automatic print functions
• Print buffer records or count averages
Note: Filtered data is printed from the Historical screen.
Set automatic print functions
Note: If the sample period is very brief and the hold time is zero, the sending of
some sample data to the printer may be skipped. However, the data is always
obtained and saved. To avoid a printout failure, set the total time of the Sample
period and the Hold period combined to more than 12 seconds.
1. On the Counter Navigation screen, push PRINTER.
2. On the Print Center screen, select the Sample Print Mode field.
Select an option for automatic printing.
Option Description
None
No data will print automatically