Option Description
Alarms Prints results at the end of the sample when a count alarm is
exceeded
Cycles Prints the results of multiples of the programmed count cycle. The
range is 1–99.
All
Prints results after each count cycle is finished
3. Set the Cycle Print Order. Options: 10–1 (oldest printed first) or
1–10 (newest printed first).
4. To show the average of the values in the printout when the samples
are taken and the data is printed, select Print Averages.
Note: The fourth value in the printout is the average of the three previous
values.
5. To round values to the nearest whole number, select Round
Averages.
6. To repeat the header data on all samples, select Repeat Sample
Headers. When not selected, the headers print only with the first
sample of a series when the cycles are set to > 1.
7. Push ENTER to confirm.
Print records manually
The buffer holds 5000 records maximum. The Print Center can print the
entire buffer or the average of count cycles obtained at each sample
point. To print records manually:
1. On the counter Navigation screen, push PRINTER.
2. On the Print Center screen, select the print option for the data. In the
Print Report field, select the regulatory reports that will be printed.
3. The data will begin to print.
• To cancel the print job, push CANCEL PRINT.
• To return to Counter Navigation, push RETURN.
18 English
Maintenance
W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance.
If the internal components must be cleaned or repaired, contact the
manufacturer.
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Clean the instrument exterior
The instrument exterior can be cleaned as needed. To avoid human
exposure to potentially dangerous chemicals, make sure to clean the
touchscreen immediately after contact with chemicals.
N O T I C E
To prevent damage, do not use an aerosol cleaner or glass cleaner on the
touchscreen. Do not leave visible moisture on the instrument or touchscreen.
Moisture can penetrate the touchscreen and damage the electronics inside.
1. Put the cap on the sample air intake nozzle to keep contamination
out.
2. Spray a mild cleaning solution on a soft cloth. Wipe the outside of the
instrument carefully.
3. Use a soft, dry cloth to wipe the touchscreen surface. If needed,
moisten the soft cloth with a mild cleaning solution.
Set the count to zero
Do this procedure after unexpectedly high particle counts. This
procedure will verify that the particle counter works correctly and will
remove residual particles.
1. Put the zero-count filter on the intake tube.
2. Turn on the unit and log in if needed.