System Contraindications And Warnings - Aerogen Solo Manuel D'utilisation Du Système

Masquer les pouces Voir aussi pour Solo:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

System Contraindications and Warnings

Contraindications
Do not use the Aerogen Solo nebuliser between the wye and a neonate
patient. The total combined volume of the Aerogen Solo nebuliser, T-piece
and/or HME may increase the dead-space to the extent that it adversely
impacts the ventilatory parameters of the neonatal patient.
Do not use the Aerogen Solo nebuliser with neonatal tracheostomised
patients. The total combined volume of the Aerogen Solo nebuliser,
T-piece and tracheostomy tube assembly may increase the dead-space
to the extent that it adversely impacts the respiratory parameters of the
neonatal patient.
System Warnings
Read and study all instructions before using the Aerogen Solo System
and accessories. Only trained medical personnel should operate the
device.
This is a single patient use device not to be used on more than one patient
to prevent cross infection.
The components and accessories of the Aerogen Solo System, as packaged,
are not sterile.
The components and accessories of the Aerogen Solo System are not made
with natural rubber latex.
Inspect all parts before use, and do not use if any parts are missing,
cracked or damaged. In case of missing parts, malfunction or damage,
contact your sales representative.
Only use physician-prescribed solutions that are approved for use with
a general purpose nebuliser. Consult drug manufacturer's instructions
regarding suitability for nebulisation.